@article{oai:repo.qst.go.jp:00043960, author = {Zheng, Jian and Yamada, Masatoshi and 鄭 建 and 山田 正俊}, journal = {Talanta}, month = {Jan}, note = {The determination of uranium is important for environmental radioactivity monitoring, which investigates the releases of uranium from nuclear facilities and of naturally occurring radioactive materials by the coal, oil, natural gas, mineral, ore refining and phosphate fertilizer industries, and it is also important for studies on the biogeochemical behavior of uranium in the environment. In this paper, we describe a quadrupole ICP-MS based analytical procedure for the accurate determination of U isotope ratios (235U/238U atom ratio and 234U/238U activity ratio) in sediment samples. A two-stage sample cleanup using anion-exchange and TEVA extraction chromatography was employed in order to obtain accurate and precise 234U/238U activity ratios. The factors that affect the accuracy and precision of U isotope ratio analysis, such as detector dead time, abundance sensitivity, dwell time and mass bias were carefully evaluated and corrected. With natural U, a precision lower than 0.5% RSD for 235U/238U atom ratio and lower than 2.0 % RSD for 234U/238U activity ratio was obtained with less than 90 ng uranium. The developed analytical method was validated using an ocean sediment reference material and applied to an investigation into the uranium isotopic compositions in a sediment core in a brackish lake in the vicinity of U-related nuclear facilities in Japan.}, pages = {932--939}, title = {Determination of U isotope ratios in sediments using ICP-QMS after sample cleanup with anion-exchange and extraction chromatography}, volume = {68}, year = {2006} }