@article{oai:repo.qst.go.jp:00043815, author = {石川, 正純 and 平澤, 雅彦 and 富谷, 武浩 and 村山, 秀雄 and 星, 正治 and 平澤 雅彦 and 富谷 武浩 and 村山 秀雄 and 星 正治}, issue = {1}, journal = {医学物理}, month = {Mar}, note = {The kinematics of boron compounds in vivo is crucial to the boron neutron capture therapy. In this paper, we proposed the prompt gamma-ray Compton scatter camera (PG-CSC), which can evaluate the 3-D kinematics of boron compounds without invasion. This PG-CSC combines a Compton camera with the prompt gamma-ray analysis. The results of its design optimization and tests with simulations, showed that one-hour-measurement was enough for evaluating a 3-D boron concentration distribution in a rat's brain with the resolution of 1mm in FWHM.}, pages = {3--12}, title = {非侵襲的ホウ素濃度分布測定システムの基礎的検討}, volume = {25}, year = {2005} }