@article{oai:repo.qst.go.jp:00043082, author = {Dachev, Tsvetan and Dimitrov, Plamen and Lemaire, Jacques and Gregoire, Ghislain and Cyamukungu, Mathias and Schmitz, H and Fujitaka, Kazunobu and Uchihori, Yukio and Kitamura, Hisashi and Reitz, Gunther and Beaujean, Rudolf and Petrov, Vladislav and Shurshakov, Vyacheslav and Benghin, Victor and 藤高 和信 and 内堀 幸夫 and 北村 尚 and ブジョーン ルドルフ}, journal = {Advances in Space Research}, month = {}, note = {The Mobile Radiation Exposure Control System's (Liulin-4 type) main purpose is to monitor simultaneously the doses and fluxes at 4 independent places. It can also be used for personnel dosimetry. The system consists of 4 battery-operated 256-channel dosimeters-spectrometers. We describe results obtained during the calibrations of the spectrometers at the Cyclotron facilities of the University of Louvain, Belgium and of the National Institute of Radiological Sciences-STA, Chiba, Japan with protons of energies up to 70 MeV. The angular sensitivities of the devices are studied and compared with Monte-Carlo predictions. We also present the results obtained at the HIMAC accelerator with 500 MeV/u Fe ions. Records made during airplane flights are shown and compared with the predictions of the CARY-3 model.}, pages = {917--925}, title = {Calibration Results Obtained With Liulin-4 Type Dosimeters}, volume = {30}, year = {2002} }