  1. 原著論文

Issues in monitoring orthopedic surgeons with active personal dosemeters when exposed to pulsed X-ray fields during orthopedic procedures

https://repo.qst.go.jp/records/2001926
https://repo.qst.go.jp/records/2001926
8be7a29d-3b2a-4194-9e05-beaae97375aa
アイテムタイプ 学術雑誌論文 / Journal Article(1)
公開日 2025-11-11
タイトル
タイトル Issues in monitoring orthopedic surgeons with active personal dosemeters when exposed to pulsed X-ray fields during orthopedic procedures
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Kowatari Munehiko

× Kowatari Munehiko

Kowatari Munehiko

Search repository
Nakagami Koichi

× Nakagami Koichi

Nakagami Koichi

Search repository
Yasukazu Hijikata

× Yasukazu Hijikata

Yasukazu Hijikata

Search repository
Kazuta Yamashita

× Kazuta Yamashita

Kazuta Yamashita

Search repository
Hirai Yuta

× Hirai Yuta

Hirai Yuta

Search repository
Matsuzaki Satoru

× Matsuzaki Satoru

Matsuzaki Satoru

Search repository
Moritake Takashi

× Moritake Takashi

Moritake Takashi

Search repository
Tomoya Tsuji

× Tomoya Tsuji

Tomoya Tsuji

Search repository
Hayo Zutz

× Hayo Zutz

Hayo Zutz

Search repository
Steffen Ketelhut

× Steffen Ketelhut

Steffen Ketelhut

Search repository
Oliver Hupe

× Oliver Hupe

Oliver Hupe

Search repository
抄録
内容記述タイプ Abstract
内容記述 It is crucial to monitor healthcare workers’ radiation exposure, especially given the increasing reliance on X-ray imaging technologies. In response to new evidence that the threshold for lens opacity is lower than previously thought, the International Commission on Radiological Protection (ICRP) has revised the annual dose limit for exposure to the lens of the eye. After the revision of the annual eye lens dose limit, the new survey on the dosemeter wearing rate was conducted in Japan. The result showed that despite the presence of monitoring protocols, the dosemeter wearing rate among healthcare workers was still low. Compliance often lacked robust verification measures, especially in smaller medical facilities. Active personal dosemeters (APDs) are commonly used in Japan, although studies suggest that they may underestimate exposure in certain settings, such as pulsed X-ray fields. Passive dosemeters, on the other hand, serve as legal dosemeters, offering a more accurate assessment of radiation exposure. This article addresses issues related to the monitoring of healthcare workers using APDs and clarifies concerns regarding the use of APDs by orthopedic surgeons during procedures involving high-dose X-ray exposure. A series of performance tests suggested that the APDs worn by orthopedic surgeons, who are often exposed to higher dose rates near patients, may underestimate radiation levels during procedures. This underestimation occurs because the surgeons are exposed to pulsed, scattered X-ray fields with energies ranging from 20 and 100 keV. The results also suggest monitoring techniques should be improved to ensure the safety of healthcare workers.
書誌情報 Radiation Protection Dosimetry

巻 201, 号 9, p. 663-671, 発行日 2025-06
DOI
識別子タイプ DOI
関連識別子 10.1093/rpd/ncaf061
戻る
0
views
See details
Views

Versions

Ver.1 2026-01-06 05:28:36.013079
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3