Radiation source detection for the accurate location of lymph node metastases using robotic forceps-type coincidence radiation detector

https://repo.qst.go.jp/records/2001396
https://repo.qst.go.jp/records/2001396
88ed10ad-53bc-4a5e-864d-66841f6145c5
アイテムタイプ 学術雑誌論文 / Journal Article(1)
公開日 2024-12-10
タイトル
タイトル Radiation source detection for the accurate location of lymph node metastases using robotic forceps-type coincidence radiation detector
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Kazuya Kawamura

× Kazuya Kawamura

Kazuya Kawamura

Search repository
Ayano Nakajima

× Ayano Nakajima

Ayano Nakajima

Search repository
Shigeki Ito

× Shigeki Ito

Shigeki Ito

Search repository
Takahashi Miwako

× Takahashi Miwako

Takahashi Miwako

Search repository
Yamaya Taiga

× Yamaya Taiga

Yamaya Taiga

Search repository
抄録
内容記述タイプ Abstract
内容記述 Purpose:
We have developed a forceps-type coincidence radiation detector for supporting lymph node dissection in esophageal cancer treatment. For precise detecting, this study aims to measure the 2D point-spread function of the detector at three
difference tip angles, to devise a method to determine the position of a point source using the 2D point-spread function.
Method: The 2D sensitivity distribution on the surface of the detector was investigated to assess sensitivity variation caused by differences in the relative positions of the detector and radiation source. Based on the results, we identified the peak sensitivity value and proposed a detection method using this value. We evaluated the effectiveness of the proposed method by detecting radiation source location using this simulated distribution.
Result :From the radiation sensitivity distribution measurements, we observed a gradual decrease in radiation detection sensitivity from the center toward the edges of the detector surface. Additionally, we verified that the peak sensitivity value was attainable. Through the basic verification of the detection method, we confirmed that the radiation source location could be detected within a maximum error of 1.4 mm.
Conclusion: We developed a peak value search method aimed at mitigating sensitivity variations by leveraging the sensitivity distribution across the detector surface. The proposed device is thought to be able to quantitatively evaluate the desired target assuming that the field of view could be limited to the area clamped by the detector. As a next research step, more precise search methods should be verified in an environment resembling the one of the target clinical uses.
書誌情報 International Journal of Computer Assisted Radiology and Surgery

発行日 2024-12
DOI
識別子タイプ DOI
関連識別子 10.1007/s11548-024-03296-8
戻る
0
views
See details
Views

Versions

Ver.1 2025-08-15 05:51:30.845752
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3