Predicting blurless and noise-free Ir-192 source images from gamma camera images for high-dose-rate brachytherapy using a deep-learning approach

https://repo.qst.go.jp/records/2000888
https://repo.qst.go.jp/records/2000888
d0c7c067-5d89-4af1-8268-e7b1a0adbd35
アイテムタイプ 学術雑誌論文 / Journal Article(1)
公開日 2025-04-21
タイトル
タイトル Predicting blurless and noise-free Ir-192 source images from gamma camera images for high-dose-rate brachytherapy using a deep-learning approach
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Kohei Nakanishi(名大)

× Kohei Nakanishi(名大)

Kohei Nakanishi(名大)

Search repository
Seiichi Yamamoto(早稲田大)

× Seiichi Yamamoto(早稲田大)

Seiichi Yamamoto(早稲田大)

Search repository
Yabe Takuya

× Yabe Takuya

Yabe Takuya

Search repository
Katsunori Yogo(名大)

× Katsunori Yogo(名大)

Katsunori Yogo(名大)

Search repository
Yumiko Noguchi(名大)

× Yumiko Noguchi(名大)

Yumiko Noguchi(名大)

Search repository
Kuniyasu Okudaira(名大)

× Kuniyasu Okudaira(名大)

Kuniyasu Okudaira(名大)

Search repository
Kawachi Naoki

× Kawachi Naoki

Kawachi Naoki

Search repository
Jun Kataoka(早稲田大)

× Jun Kataoka(早稲田大)

Jun Kataoka(早稲田大)

Search repository
抄録
内容記述タイプ Abstract
内容記述 Objective: Precise monitoring of the position and dwell time of iridium-192 (Ir-192) during high-dose- rate (HDR) brachytherapy is crucial to avoid serious damage to normal tissues. Source imaging using a compact gamma camera is a potential approach for monitoring. However, images from the gamma camera are affected by blurring and statistical noise, which impact the accuracy of source position monitoring. This study aimed to develop a deep-learning approach for estimating ideal source images that reduce the effect of blurring and statistical noise from experimental images captured using a compact gamma camera. Approach: A double pix2pix model was trained using the simulated gamma camera images of an Ir-192 source. The first model was responsible for denoising the Ir-192 images, whereas the second model performed super resolution. Trained models were then applied to the experimental images to estimate the ideal images.
Main results: At a distance of 100 mm between the compact gamma camera and the Ir-192 source, the difference in full width at half maximum (FWHM) between the estimated and actual source sizes was approximately 0.5 mm for a measurement time of 1.5 s. This difference has been improved from approximately 2.7 mm without the use of DL. Even with a measurement time of 0.1 s, the ideal images could be estimated as accurately as in the 1.5 s measurements. This method consistently achieved accurate estimations of the source images at any position within the field of view; however, the difference increased with the distance between the Ir-192 source and the compact gamma camera.
Significance: The proposed method successfully provided estimated images from the experimental images within errors smaller than 0.5 mm at 100 mm. This method is promising for reducing blurring and statistical noise from the experimental images, enabling precise real-time monitoring of Ir-192 sources during HDR brachytherapy.
書誌情報 Biomedical Physics & Engineering Express

巻 10, 号 1, p. 015006, 発行日 2023-11
出版者
出版者 IOP Publishing
ISSN
収録物識別子タイプ ISSN
収録物識別子 2057-1976
DOI
識別子タイプ DOI
関連識別子 10.1088/2057-1976/ad0bb2
戻る
0
views
See details
Views

Versions

Ver.1 2025-08-15 03:04:21.660491
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3