| アイテムタイプ |
学術雑誌論文 / Journal Article(1) |
| 公開日 |
2025-01-08 |
| タイトル |
|
|
タイトル |
Standardization of Rn-222 concentration using the multi-electrode proportional counter |
|
言語 |
en |
| 言語 |
|
|
言語 |
eng |
| 資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_6501 |
|
資源タイプ |
journal article |
| 著者 |
Rio Furukawa
Miroslaw Janik
Kodaira Satoshi
Seiya Manabe
Tetsuro Matsumoto
Chihiro Shimodan
Yasushi Sato
Hideki Harano
|
| 抄録 |
|
|
内容記述タイプ |
Abstract |
|
内容記述 |
Standardization of the concentration of gaseous 222Rn based on a multi-electrode proportional counter (MEPC) is under development as a primary standard in Japan. In this study, the concept and evaluation of its performance are reported. The latter consists of a preliminary result for the uncertainty budget associated with the measurement of the MEPC and compensation of the electric field distortion in the MEPC. Moreover, an ionization-chamber-based gas circulation system was added for the calibration of radon monitors in the air. |
| 書誌情報 |
Applied Radiation and Isotopes
巻 202,
p. 111076,
発行日 2023-10
|
| 出版者 |
|
|
出版者 |
ELSEVIER |
| ISSN |
|
|
収録物識別子タイプ |
ISSN |
|
収録物識別子 |
0969-8043 |
| DOI |
|
|
|
識別子タイプ |
DOI |
|
|
関連識別子 |
10.1016/j.apradiso.2023.111076 |