| アイテムタイプ |
学術雑誌論文 / Journal Article(1) |
| 公開日 |
2022-10-28 |
| タイトル |
|
|
タイトル |
Two-dimensional positronium lifetime imaging using certified reference materials |
|
言語 |
en |
| 言語 |
|
|
言語 |
eng |
| 資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_6501 |
|
資源タイプ |
journal article |
| 著者 |
Takyu Sodai
Shibuya Kengo
Nishikido Fumihiko
Tashima Hideaki
Takahashi Miwako
Yamaya Taiga
|
| 抄録 |
|
|
内容記述タイプ |
Abstract |
|
内容記述 |
Positronium (Ps) is generated before electron?positron annihilation in positron emission tomography (PET). As the Ps lifetime reflects the surrounding electron density, we aimed at its use for an innovative PET imaging. Here, as our first step, we demonstrated two-dimensional Ps lifetime imaging of certified reference materials using a pair of time-of-flight PET detectors. The imaged and calculated Ps lifetimes were very close to those of the reference materials, but the error values were larger than those of the reference materials. The results suggested that the accuracy of PET imaging with the ortho-positronium lifetime could be verified by those materials. |
| 書誌情報 |
Applied Physics Express
巻 15,
号 10,
p. 106001,
発行日 2022-09
|
| 出版者 |
|
|
出版者 |
The Japan Society of Applied Physics |
| ISSN |
|
|
収録物識別子タイプ |
ISSN |
|
収録物識別子 |
1882-0778 |
| DOI |
|
|
|
識別子タイプ |
DOI |
|
|
関連識別子 |
10.35848/1882-0786/ac8d7b |