量研学術機関リポジトリ「QST-Repository」は、国立研究開発法人 量子科学技術研究開発機構に所属する職員等が生み出した学術成果(学会誌発表論文、学会発表、研究開発報告書、特許等)を集積しインターネット上で広く公開するサービスです。 Welcome to QST-Repository where we accumulates and discloses the academic research results(Journal Publications, Conference presentation, Research and Development Report, Patent, etc.) of the members of National Institutes for Quantum Science and Technology.
Thank you very much for using our website. On the 11th of March 2019, this site was moved from our own network server to the JAIRO Cloud network server. If you previously bookmarked this site, that bookmark will no longer work. We would be grateful if you could bookmark the website again. Thank you very much for your understanding and cooperation.
Particle therapy using protons or carbon ions for cancer patients with cardiac implantable electronic devices (CIED): a retrospective multi‑institutional study
利用統計を見る
Purpose
To evaluate the outcomes of particle therapy in cancer patients with cardiac implantable electronic devices (CIEDs). Materials and methods From April 2001 to March 2013, 19,585 patients were treated with proton beam therapy (PBT) or
carbon ion therapy (CIT) at 8 institutions. Of these, 69 patients (0.4%, PBT 46, CIT 22, and PBT + CIT 1) with CIEDs (64 pacemakers, 4 implantable cardioverter defibrillators, and 1 with a cardiac resynchronization therapy defibrillator) were retrospectively reviewed. All the patients with CIEDs in this study were treated with the passive scattering type of particle beam therapy.
Results
Six (13%) of the 47 PBT patients, and none of the 23 CIT patients experienced CIED malfunctions (p = 0.105). Electrical resets (7) and over-sensing (3) occurred transiently in 6 patients. The distance between the edge of the irradiation field and the CIED was not associated with the incidence of malfunctions in 20 patients with lung cancer. A larger field size had a higher event rate but the test to evaluate trends as not statistically significant (p = 0.196).
Conclusion
Differences in the frequency of occurrence of device malfunctions for patients treated with PBT and patients treated with CIT did not reach statistical significance. The present study can be regarded as a benchmark study about the incidence of malfunctioning of CIED in passive scattering particle beam therapy and can be used as a reference for active scanning particle beam therapy.