Time-resolved dosimetry with pencil-beam scanning for quality assurance/quality control in particle therapy

https://repo.qst.go.jp/records/84666
https://repo.qst.go.jp/records/84666
d4b98234-8baf-4b1c-aa0a-5c7a553f2df3
Item type 学術雑誌論文 / Journal Article(1)
公開日 2021-10-20
タイトル
タイトル Time-resolved dosimetry with pencil-beam scanning for quality assurance/quality control in particle therapy
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Han, Soorim

× Han, Soorim

WEKO 1021855

Han, Soorim

Search repository
Takuji, Furukawa

× Takuji, Furukawa

WEKO 1021856

Takuji, Furukawa

Search repository
Yousuke, Hara

× Yousuke, Hara

WEKO 1021857

Yousuke, Hara

Search repository
Shigekazu, Fukuda

× Shigekazu, Fukuda

WEKO 1021858

Shigekazu, Fukuda

Search repository
Han, Soorim

× Han, Soorim

WEKO 1021859

en Han, Soorim

Search repository
Takuji, Furukawa

× Takuji, Furukawa

WEKO 1021860

en Takuji, Furukawa

Search repository
Yousuke, Hara

× Yousuke, Hara

WEKO 1021861

en Yousuke, Hara

Search repository
Shigekazu, Fukuda

× Shigekazu, Fukuda

WEKO 1021862

en Shigekazu, Fukuda

Search repository
抄録
内容記述タイプ Abstract
内容記述 This study aimed to measure dose in a scanning carbon beam-irradiation field with high sampling rate that is sufficient for identifying spots and verifying the characteristics of the scanning beam that cannot generally be derived from the
dose. To identify the spot, which is the smallest control unit of beam information during irradiation, effecting measurements with a sampling time of 10 μs or shorter is necessary. The provided dose within a specific time is referred to as time-resolved dose (TRD).We designed a circuit for time-resolved dosimetry using a fast-data acquisition unit (SL1000, Yokogawa Electric Co.), which can measure 100 000 samples per second. Moreover, we used converters to enable a connection between an ionization chamber (IC) and the SL1000.TRD was measured successfully using point irradiation and two-dimensional irradiation patterns in a scanned carbon beam. Based on the moving time of the spot obtained from the position monitor, the dose delivered to the IC from each spot position (spot dose) was interpreted. The spot dose, displacement of the chamber from the beam’s center axis, and beam size were derived using TRD and position monitor outputs, which were measured concurrent with TRD. Spot dose up to a radius of 8 mm area from the IC’s center were observed. Using the spotdose equations and simulation, we show that the spot dose of each position varies depending on the beam size and displacement of the IC’s center from the beam’s center axis.We devise an interpretation method for the characteristics that may apply to quality assurance, such as the verification of the trend for the beam axis and isocenter to coincide, as well as beam-size verification.
書誌情報 Journal of Applied Clinical Medical Physics

巻 22, 号 11, p. 12-20, 発行日 2021-11
出版者
出版者 AAPM
ISSN
収録物識別子タイプ ISSN
収録物識別子 1526-9914
DOI
識別子タイプ DOI
関連識別子 10.1002/acm2.13397
関連サイト
識別子タイプ URI
関連識別子 https://aapm.onlinelibrary.wiley.com/doi/full/10.1002/acm2.13397
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 17:08:42.756102
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3