WEKO3
アイテム
Characteristics of fluorescent nuclear track detection with Ag+-activated phosphate glass
https://repo.qst.go.jp/records/78565
https://repo.qst.go.jp/records/78565183755d5-ab86-4b77-8626-2bb90182fbf9
| Item type | 学術雑誌論文 / Journal Article(1) | |||||
|---|---|---|---|---|---|---|
| 公開日 | 2020-01-29 | |||||
| タイトル | ||||||
| タイトル | Characteristics of fluorescent nuclear track detection with Ag+-activated phosphate glass | |||||
| 言語 | ||||||
| 言語 | eng | |||||
| 資源タイプ | ||||||
| 資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
| 資源タイプ | journal article | |||||
| アクセス権 | ||||||
| アクセス権 | metadata only access | |||||
| アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
| 著者 |
Kodaira, Satoshi
× Kodaira, Satoshi× Kusumoto, Tamon× Kitamura, Hisashi× Yanagida, Yuka× Koguchi, Yasuhiro× Satoshi, Kodaira× Tamon, Kusumoto× Hisashi, Kitamura |
|||||
| 抄録 | ||||||
| 内容記述タイプ | Abstract | |||||
| 内容記述 | Silver-activated phosphate glass (Ag+-glass), which is commonly employed as a radiophotoluminscence (RPL) dosimeter, was applied as a fluorescent nuclear track detector (FNTD). The heavy ions from C to Fe ions were detected as fluorescent nuclear tracks under a confocal microscope. The detection linear energy transfer (LET) threshold was estimated to be 5 keV/μm in water. The fading effect was observed in which the fluorescent intensity decreases with repeated laser scans. The Ag+-glass has good potential applicability as the FNTD for the high LET range (5– 200 keV/μm) in radioimmunotherapy and space radiation dosimetry uses. | |||||
| 書誌情報 |
Radiation Measurements 巻 132, p. 106252, 発行日 2020-03 |
|||||
| 出版者 | ||||||
| 出版者 | Elsevier | |||||
| ISSN | ||||||
| 収録物識別子タイプ | ISSN | |||||
| 収録物識別子 | 1350-4487 | |||||
| DOI | ||||||
| 識別子タイプ | DOI | |||||
| 関連識別子 | 10.1016/j.radmeas.2020.106252 | |||||
| 関連サイト | ||||||
| 識別子タイプ | DOI | |||||
| 関連識別子 | https://doi.org/10.1016/j.radmeas.2020.106252 | |||||