Compton imaging with 99mTc for human imaging

https://repo.qst.go.jp/records/76770
https://repo.qst.go.jp/records/76770
e65c865a-3812-4afd-88e8-c7d3044e4f40
Item type 学術雑誌論文 / Journal Article(1)
公開日 2019-09-12
タイトル
タイトル Compton imaging with 99mTc for human imaging
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Sakai, Makoto

× Sakai, Makoto

WEKO 846259

Sakai, Makoto

Search repository
Kubota, Yoshiki

× Kubota, Yoshiki

WEKO 846260

Kubota, Yoshiki

Search repository
Kumar Parajuli, Raj

× Kumar Parajuli, Raj

WEKO 846261

Kumar Parajuli, Raj

Search repository
Kikuchi, Mikiko

× Kikuchi, Mikiko

WEKO 846262

Kikuchi, Mikiko

Search repository
Arakawa, Kazuo

× Arakawa, Kazuo

WEKO 846263

Arakawa, Kazuo

Search repository
Nakano, Takashi

× Nakano, Takashi

WEKO 846264

Nakano, Takashi

Search repository
Parajuli, Raj

× Parajuli, Raj

WEKO 846265

en Parajuli, Raj

Search repository
Nakano, Takashi

× Nakano, Takashi

WEKO 846266

en Nakano, Takashi

Search repository
抄録
内容記述タイプ Abstract
内容記述 We have been developing a medical imaging system using a Compton camera and demonstrated the imaging ability of Compton camera for Tc-DMSA accumulated in rat kidneys. In this study, we performed imaging experiments using a human body phantom to confirm its applicability to human imaging. Preliminary simulations were conducted using a digital phantom with varying activity ratios between the kidney and body trunk regions. Gamma rays (141 keV) were generated and detected by a Compton camera based on a silicon and cadmium telluride (Si/CdTe) detector. Compton images were reconstructed with the list mode median root prior expectation maximization method. The appropriate number of iterations of the condition was confirmed through simulations. The reconstructed Compton images revealed two bright points in the kidney regions. Furthermore, the numerical value calculated by integrating pixel values inside the region of interest correlated well with the activity of the kidney regions. Finally, experimental studies were conducted to ascertain whether the results of the simulation studies could be reproduced. The kidneys could be successfully visualised. In conclusion, considering that the conditions in this study agree with those of typical human bodies and imaginable experimental setup, the Si/CdTe Compton camera has a high probability of success in human imaging. In addition, our results indicate the capability of (semi-) quantitative analysis using Compton images.
書誌情報 Scientific reports

巻 9, 号 1, p. 12906, 発行日 2019-09
出版者
出版者 Nature
ISSN
収録物識別子タイプ ISSN
収録物識別子 2045-2322
PubMed番号
識別子タイプ PMID
関連識別子 31501461
DOI
識別子タイプ DOI
関連識別子 10.1038/s41598-019-49130-z
関連サイト
識別子タイプ URI
関連識別子 https://www.nature.com/articles/s41598-019-49130-z
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 18:18:27.795792
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3