{"_buckets": {"deposit": "04689963-006c-41f4-9462-1a2da448cc18"}, "_deposit": {"created_by": 1, "id": "72422", "owners": [1], "pid": {"revision_id": 0, "type": "depid", "value": "72422"}, "status": "published"}, "_oai": {"id": "oai:repo.qst.go.jp:00072422", "sets": ["28"]}, "author_link": ["713344", "713341", "713348", "713337", "713350", "713339", "713336", "713340", "713338", "713347", "713343", "713349", "713345", "713342", "713346", "713335"], "item_10005_date_7": {"attribute_name": "発表年月日", "attribute_value_mlt": [{"subitem_date_issued_datetime": "2017-07-30", "subitem_date_issued_type": "Issued"}]}, "item_10005_description_5": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "Purpose: At the National Institute of Radiological Sciences (NIRS), we have been commissioning the rotating- gantry system for the carbon-ion radiotherapy in 2015. It is world’s first rotating-gantry with super-conducting magnet and achieved a more compact, lighter rotating gantry than the typical gantry. During the commissioning of the rotating gantry, we adjusted the beam position using a steel sphere located at isocenter and verified the accuracy using a scintillation screen and CCD system. We employ the digital QA system to verify the geometrical accuracy easily and quickly.\n\\nMethods: The 300 ton rotating gantry can deliver carbon ions having 430 MeV/u to the isocenter over 360 degrees, having the capability of performing three-dimensional scanning irradiation. the mechanical isocenter was verified using the laser tracker system. To verify the coincidence of the isocenter for irradiation system and the patient positioning system, we used a steel sphere located at isocenter. As a part of the verifying of the treatment beam position, scintillation screen viewed by a CCD camera mounted on the gantry downstream of the sphere. Total 201energy carbon beams with that were in the range from 48 to 430 MeV/u were measured sequentially. A starshot test was also performed to verify the gantry angle and beam direction visually using a scintillator block and a CCD camera. For the starshot test, 13 beam energy were selected. \n\\nResults: From the result of the test using a sphere, the maximum displacement of the beam position was within 0.5mm at all gantry angles. The gantry starshot shows accuracy of the gantry angle was less than 0.5 degree. \n\\nConclusion: The isocentricity of the compact super-conducting rotating gantry for carbon ion therapy dedicated to the highly precise treatment were within 1mm. Two of digital measurement system brought high-efficient beam adjusting and commissioning.", "subitem_description_type": "Abstract"}]}, "item_10005_description_6": {"attribute_name": "会議概要(会議名, 開催地, 会期, 主催者等)", "attribute_value_mlt": [{"subitem_description": "59th AnnualMeeting of American association of Physicists in Medicine (AAPM2017)", "subitem_description_type": "Other"}]}, "item_access_right": {"attribute_name": "アクセス権", "attribute_value_mlt": [{"subitem_access_right": "metadata only access", "subitem_access_right_uri": "http://purl.org/coar/access_right/c_14cb"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Saotome, Naoya"}], "nameIdentifiers": [{"nameIdentifier": "713335", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Furukawa, Takuji"}], "nameIdentifiers": [{"nameIdentifier": "713336", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Saraya, Yuichi"}], "nameIdentifiers": [{"nameIdentifier": "713337", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Mizushima, Kota"}], "nameIdentifiers": [{"nameIdentifier": "713338", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Hara, Yousuke"}], "nameIdentifiers": [{"nameIdentifier": "713339", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Tansho, Ryohei"}], "nameIdentifiers": [{"nameIdentifier": "713340", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Shirai, Toshiyuki"}], "nameIdentifiers": [{"nameIdentifier": "713341", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Noda, Koji"}], "nameIdentifiers": [{"nameIdentifier": "713342", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "早乙女 直也", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "713343", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "古川 卓司", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "713344", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "皿谷 有一", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "713345", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "水島 康太", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "713346", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "原 洋介", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "713347", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "丹正 亮平", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "713348", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "白井 敏之", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "713349", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "野田 耕司", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "713350", "nameIdentifierScheme": "WEKO"}]}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "conference object", "resourceuri": "http://purl.org/coar/resource_type/c_c94f"}]}, "item_title": "Isocenter Verification of a Compact Rotating Gantry for Carbon Ion Therapy", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Isocenter Verification of a Compact Rotating Gantry for Carbon Ion Therapy"}]}, "item_type_id": "10005", "owner": "1", "path": ["28"], "permalink_uri": "https://repo.qst.go.jp/records/72422", "pubdate": {"attribute_name": "公開日", "attribute_value": "2017-08-25"}, "publish_date": "2017-08-25", "publish_status": "0", "recid": "72422", "relation": {}, "relation_version_is_last": true, "title": ["Isocenter Verification of a Compact Rotating Gantry for Carbon Ion Therapy"], "weko_shared_id": -1}
Isocenter Verification of a Compact Rotating Gantry for Carbon Ion Therapy
https://repo.qst.go.jp/records/72422
https://repo.qst.go.jp/records/72422dad2cfe7-6f30-4524-9e15-011e495241dd
Item type | 会議発表用資料 / Presentation(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-08-25 | |||||
タイトル | ||||||
タイトル | Isocenter Verification of a Compact Rotating Gantry for Carbon Ion Therapy | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
資源タイプ | conference object | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Saotome, Naoya
× Saotome, Naoya× Furukawa, Takuji× Saraya, Yuichi× Mizushima, Kota× Hara, Yousuke× Tansho, Ryohei× Shirai, Toshiyuki× Noda, Koji× 早乙女 直也× 古川 卓司× 皿谷 有一× 水島 康太× 原 洋介× 丹正 亮平× 白井 敏之× 野田 耕司 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Purpose: At the National Institute of Radiological Sciences (NIRS), we have been commissioning the rotating- gantry system for the carbon-ion radiotherapy in 2015. It is world’s first rotating-gantry with super-conducting magnet and achieved a more compact, lighter rotating gantry than the typical gantry. During the commissioning of the rotating gantry, we adjusted the beam position using a steel sphere located at isocenter and verified the accuracy using a scintillation screen and CCD system. We employ the digital QA system to verify the geometrical accuracy easily and quickly. \nMethods: The 300 ton rotating gantry can deliver carbon ions having 430 MeV/u to the isocenter over 360 degrees, having the capability of performing three-dimensional scanning irradiation. the mechanical isocenter was verified using the laser tracker system. To verify the coincidence of the isocenter for irradiation system and the patient positioning system, we used a steel sphere located at isocenter. As a part of the verifying of the treatment beam position, scintillation screen viewed by a CCD camera mounted on the gantry downstream of the sphere. Total 201energy carbon beams with that were in the range from 48 to 430 MeV/u were measured sequentially. A starshot test was also performed to verify the gantry angle and beam direction visually using a scintillator block and a CCD camera. For the starshot test, 13 beam energy were selected. \nResults: From the result of the test using a sphere, the maximum displacement of the beam position was within 0.5mm at all gantry angles. The gantry starshot shows accuracy of the gantry angle was less than 0.5 degree. \nConclusion: The isocentricity of the compact super-conducting rotating gantry for carbon ion therapy dedicated to the highly precise treatment were within 1mm. Two of digital measurement system brought high-efficient beam adjusting and commissioning. |
|||||
会議概要(会議名, 開催地, 会期, 主催者等) | ||||||
内容記述タイプ | Other | |||||
内容記述 | 59th AnnualMeeting of American association of Physicists in Medicine (AAPM2017) | |||||
発表年月日 | ||||||
日付 | 2017-07-30 | |||||
日付タイプ | Issued |