{"_buckets": {"deposit": "0c4d0468-0cd3-4421-a440-97a2d4716d8c"}, "_deposit": {"created_by": 1, "id": "54348", "owners": [1], "pid": {"revision_id": 0, "type": "depid", "value": "54348"}, "status": "published"}, "_oai": {"id": "oai:repo.qst.go.jp:00054348", "sets": ["2"]}, "author_link": ["555177", "555174", "555179", "555180", "555176", "555175", "555178"], "item_10003_biblio_info_7": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2012-10", "bibliographicIssueDateType": "Issued"}, "bibliographicPageEnd": "81", "bibliographicPageStart": "79", "bibliographicVolumeNumber": "3", "bibliographic_titles": [{"bibliographic_title": "Progress in Nuclear Science and Technology"}]}]}, "item_10003_description_5": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "Aluminum Oxide is well known as a TLD material for environmental dosemeasurement. The use of aluminum oxide TLD became popular due to its high sensitivity, small fading, and physical stability. However the TL material was not easy to analyze due to thermal quenching and light fading effect. The development of Optically Stimulated Luminescence (OSL) method made Aluminum Oxide simplified the analytical process and equipment. A light emitting diode (LED) stimulation method was developed, which enabled the development of a compact reader, called microStar. This study was conducted to understand the characteristic of low dose measurement using the microStar and its application for environmental dose measurement. OSL dosimeters (OSLD) were irradiated using a 74 GBq and 7.4 GBq Cs-137 radiation source at National Institute of Radiological Science. Irradiations included various delivered doses from 20 KGy to 1 mGy free in air. The microStar gets good read out results. The microStar can be used for environmental dose measurement. Over 10 microStar readers have been deployed around the Fukushima Daiichi Nuclear Power Plant (NPP) for monitoring members of the general public, workers, and environmental dose.", "subitem_description_type": "Abstract"}]}, "item_10003_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "2185-4823", "subitem_source_identifier_type": "ISSN"}]}, "item_access_right": {"attribute_name": "アクセス権", "attribute_value_mlt": [{"subitem_access_right": "metadata only access", "subitem_access_right_uri": "http://purl.org/coar/access_right/c_14cb"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Kobayashi, Ikuo"}], "nameIdentifiers": [{"nameIdentifier": "555174", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Okazaki, Tohoru"}], "nameIdentifiers": [{"nameIdentifier": "555175", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Yajima, Kazuaki"}], "nameIdentifiers": [{"nameIdentifier": "555176", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Yasuda, Hiroshi"}], "nameIdentifiers": [{"nameIdentifier": "555177", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "小林 育夫", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "555178", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "矢島 千秋", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "555179", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "保田 浩志", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "555180", "nameIdentifierScheme": "WEKO"}]}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "conference paper", "resourceuri": "http://purl.org/coar/resource_type/c_5794"}]}, "item_title": "Environmental radiation dosimetry by the small OSL reader", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Environmental radiation dosimetry by the small OSL reader"}]}, "item_type_id": "10003", "owner": "1", "path": ["2"], "permalink_uri": "https://repo.qst.go.jp/records/54348", "pubdate": {"attribute_name": "公開日", "attribute_value": "2012-12-19"}, "publish_date": "2012-12-19", "publish_status": "0", "recid": "54348", "relation": {}, "relation_version_is_last": true, "title": ["Environmental radiation dosimetry by the small OSL reader"], "weko_shared_id": -1}
Environmental radiation dosimetry by the small OSL reader
https://repo.qst.go.jp/records/54348
https://repo.qst.go.jp/records/543487f5cc1d9-10ce-402e-b1fd-a293c6b2b401
Item type | 会議発表論文 / Conference Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2012-12-19 | |||||
タイトル | ||||||
タイトル | Environmental radiation dosimetry by the small OSL reader | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_5794 | |||||
資源タイプ | conference paper | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Kobayashi, Ikuo
× Kobayashi, Ikuo× Okazaki, Tohoru× Yajima, Kazuaki× Yasuda, Hiroshi× 小林 育夫× 矢島 千秋× 保田 浩志 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Aluminum Oxide is well known as a TLD material for environmental dosemeasurement. The use of aluminum oxide TLD became popular due to its high sensitivity, small fading, and physical stability. However the TL material was not easy to analyze due to thermal quenching and light fading effect. The development of Optically Stimulated Luminescence (OSL) method made Aluminum Oxide simplified the analytical process and equipment. A light emitting diode (LED) stimulation method was developed, which enabled the development of a compact reader, called microStar. This study was conducted to understand the characteristic of low dose measurement using the microStar and its application for environmental dose measurement. OSL dosimeters (OSLD) were irradiated using a 74 GBq and 7.4 GBq Cs-137 radiation source at National Institute of Radiological Science. Irradiations included various delivered doses from 20 KGy to 1 mGy free in air. The microStar gets good read out results. The microStar can be used for environmental dose measurement. Over 10 microStar readers have been deployed around the Fukushima Daiichi Nuclear Power Plant (NPP) for monitoring members of the general public, workers, and environmental dose. | |||||
書誌情報 |
Progress in Nuclear Science and Technology 巻 3, p. 79-81, 発行日 2012-10 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 2185-4823 |