An intercomparison done at NIRS, Japan on continuous monitors for measuring 220Rn concentration

https://repo.qst.go.jp/records/47605
https://repo.qst.go.jp/records/47605
b7f956e5-669a-49c7-90d3-5f3e88e1ccfc
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-01-06
タイトル
タイトル An intercomparison done at NIRS, Japan on continuous monitors for measuring 220Rn concentration
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Sorimachi, Atsuyuki

× Sorimachi, Atsuyuki

WEKO 477181

Sorimachi, Atsuyuki

Search repository
ミロソラフ, ヤニック

× ミロソラフ, ヤニック

WEKO 477182

ミロソラフ, ヤニック

Search repository
Tokonami, Shinji

× Tokonami, Shinji

WEKO 477183

Tokonami, Shinji

Search repository
Ishikawa, Tetsuo

× Ishikawa, Tetsuo

WEKO 477184

Ishikawa, Tetsuo

Search repository
ミロソラフ ヤニック

× ミロソラフ ヤニック

WEKO 477185

en ミロソラフ ヤニック

Search repository
抄録
内容記述タイプ Abstract
内容記述 An intercomparison for continuous monitors that measure thoron (220Rn) concentration was carried out
using a 220Rn chamber of National Institute of Radiological Sciences (NIRS), Japan; eleven 220Rn monitors
(four types) from nine laboratories were evaluated. The 220Rn detection principle was the same for the
eleven instruments and one reference instrument, which were commercially available silicon semiconductor detectors using an electrostatic collection method. The intercomparison results showed that there was a negative deviation of more than 30% in measured 220Rn concentrations given by the laboratories relative to the reference values, which were obtained by making a decay-correction during the travel of 220Rn through the sampling assembly (sample tube, filter and drying unit) and using a calibration factor. In order to elucidate the reason for this and then to investigate factors that affect the 220Rn concentration measured with the monitors. As a result, it was necessary to make the decay-correction, in particular, when using a drying unit with a large inner volume and to use the calibration factor in order to better estimate the 220Rn concentration. It was also found to be better to determine a calibration factor inherent to an individual monitor, because the calibration factor values ranged from 0.75 to 2.32, depending on the flow rate of the monitor pump (0.37–1.02 L min1). It was concluded from this study that a periodical calibration of the monitor and a check of the monitor flow rate during measurements are necessary to maintain a consistent quality level of the 220Rn measurement.
書誌情報 Applied Radiation and Isotopes

巻 107, p. 145-151, 発行日 2016-01
出版者
出版者 Elsevier
ISSN
収録物識別子タイプ ISSN
収録物識別子 0969-8043
DOI
識別子タイプ DOI
関連識別子 doi:10.1016/j.apradiso.2015.10.007
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 23:37:41.781652
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3