{"_buckets": {"deposit": "cdeeb384-81d4-4476-aa71-213a7bac8387"}, "_deposit": {"created_by": 1, "id": "46901", "owners": [1], "pid": {"revision_id": 0, "type": "depid", "value": "46901"}, "status": "published"}, "_oai": {"id": "oai:repo.qst.go.jp:00046901", "sets": ["1"]}, "author_link": ["468021", "468032", "468023", "468031", "468030", "468029", "468028", "468026", "468027", "468033", "468024", "468022", "468025"], "item_8_biblio_info_7": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2013-10", "bibliographicIssueDateType": "Issued"}, "bibliographicIssueNumber": "7", "bibliographicPageEnd": "1450", "bibliographicPageStart": "1446", "bibliographicVolumeNumber": "63", "bibliographic_titles": [{"bibliographic_title": "Journal of the Korean Physical Society"}]}]}, "item_8_description_5": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "A two-dimensional dosimetry system was developed for the quality assurance (QA) of therapeutic scanned ion beams at Heavy Ion Medical Accelerator in Chiba (HIMAC). This system consists of a fluorescent screen and a charge-coupled device (CCD) camera. To evaluate the performance of this system, we carried out a few experiments concerning QA procedures. A verification of this system\nwas also carried out by comparing the film dosimetry and the ionization chamber array. Also, we verified several types of corrections for errors, e.g., background and vignetting, that distort the measurement results. As a result, we confirmed that this system could be used for QA procedures of the therapeutic scanned ion beams.", "subitem_description_type": "Abstract"}]}, "item_8_publisher_8": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "Springer Netherlands"}]}, "item_8_relation_14": {"attribute_name": "DOI", "attribute_value_mlt": [{"subitem_relation_type_id": {"subitem_relation_type_id_text": "10.3938/jkps.63.1446", "subitem_relation_type_select": "DOI"}}]}, "item_8_relation_17": {"attribute_name": "関連サイト", "attribute_value_mlt": [{"subitem_relation_name": [{"subitem_relation_name_text": "http://link.springer.com/article/10.3938%2Fjkps.63.1446"}], "subitem_relation_type_id": {"subitem_relation_type_id_text": "http://link.springer.com/article/10.3938%2Fjkps.63.1446", "subitem_relation_type_select": "URI"}}]}, "item_8_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "0374-4884", "subitem_source_identifier_type": "ISSN"}]}, "item_access_right": {"attribute_name": "アクセス権", "attribute_value_mlt": [{"subitem_access_right": "metadata only access", "subitem_access_right_uri": "http://purl.org/coar/access_right/c_14cb"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Hara, Yousuke"}], "nameIdentifiers": [{"nameIdentifier": "468021", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Furukawa, Takuji"}], "nameIdentifiers": [{"nameIdentifier": "468022", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Takeshita, Eri"}], "nameIdentifiers": [{"nameIdentifier": "468023", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Mizushima, Kota"}], "nameIdentifiers": [{"nameIdentifier": "468024", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Satou, Shinji"}], "nameIdentifiers": [{"nameIdentifier": "468025", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Shirai, Toshiyuki"}], "nameIdentifiers": [{"nameIdentifier": "468026", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Noda, Kouji"}], "nameIdentifiers": [{"nameIdentifier": "468027", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "原 洋介", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "468028", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "古川 卓司", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "468029", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "水島 康太", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "468030", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "佐藤 眞二", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "468031", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "白井 敏之", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "468032", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "野田 耕司", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "468033", "nameIdentifierScheme": "WEKO"}]}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "journal article", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "Performance of a fluorescent screen with a CCD system for quality assurance in a heavy-ion beam scanning irradiation system", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Performance of a fluorescent screen with a CCD system for quality assurance in a heavy-ion beam scanning irradiation system"}]}, "item_type_id": "8", "owner": "1", "path": ["1"], "permalink_uri": "https://repo.qst.go.jp/records/46901", "pubdate": {"attribute_name": "公開日", "attribute_value": "2014-11-06"}, "publish_date": "2014-11-06", "publish_status": "0", "recid": "46901", "relation": {}, "relation_version_is_last": true, "title": ["Performance of a fluorescent screen with a CCD system for quality assurance in a heavy-ion beam scanning irradiation system"], "weko_shared_id": -1}
Performance of a fluorescent screen with a CCD system for quality assurance in a heavy-ion beam scanning irradiation system
https://repo.qst.go.jp/records/46901
https://repo.qst.go.jp/records/4690149bdedab-1423-4e6d-b08a-e49d344b12f7
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2014-11-06 | |||||
タイトル | ||||||
タイトル | Performance of a fluorescent screen with a CCD system for quality assurance in a heavy-ion beam scanning irradiation system | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Hara, Yousuke
× Hara, Yousuke× Furukawa, Takuji× Takeshita, Eri× Mizushima, Kota× Satou, Shinji× Shirai, Toshiyuki× Noda, Kouji× 原 洋介× 古川 卓司× 水島 康太× 佐藤 眞二× 白井 敏之× 野田 耕司 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A two-dimensional dosimetry system was developed for the quality assurance (QA) of therapeutic scanned ion beams at Heavy Ion Medical Accelerator in Chiba (HIMAC). This system consists of a fluorescent screen and a charge-coupled device (CCD) camera. To evaluate the performance of this system, we carried out a few experiments concerning QA procedures. A verification of this system was also carried out by comparing the film dosimetry and the ionization chamber array. Also, we verified several types of corrections for errors, e.g., background and vignetting, that distort the measurement results. As a result, we confirmed that this system could be used for QA procedures of the therapeutic scanned ion beams. |
|||||
書誌情報 |
Journal of the Korean Physical Society 巻 63, 号 7, p. 1446-1450, 発行日 2013-10 |
|||||
出版者 | ||||||
出版者 | Springer Netherlands | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0374-4884 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.3938/jkps.63.1446 | |||||
関連サイト | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://link.springer.com/article/10.3938%2Fjkps.63.1446 | |||||
関連名称 | http://link.springer.com/article/10.3938%2Fjkps.63.1446 |