Positional Dependence of the CT number use of a Cone-beam CT Scanner for an Electron Density Phantom in Particle Beam Therapy

https://repo.qst.go.jp/records/46682
https://repo.qst.go.jp/records/46682
5f90a7c1-a5dc-498a-a296-7120ae59c7bf
Item type 学術雑誌論文 / Journal Article(1)
公開日 2013-12-12
タイトル
タイトル Positional Dependence of the CT number use of a Cone-beam CT Scanner for an Electron Density Phantom in Particle Beam Therapy
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Kusano, Yohsuke

× Kusano, Yohsuke

WEKO 465460

Kusano, Yohsuke

Search repository
Uesaka, Saki

× Uesaka, Saki

WEKO 465461

Uesaka, Saki

Search repository
Yajima, Kaori

× Yajima, Kaori

WEKO 465462

Yajima, Kaori

Search repository
Kumagai, Motoki

× Kumagai, Motoki

WEKO 465463

Kumagai, Motoki

Search repository
Mizuno, Hideyuki

× Mizuno, Hideyuki

WEKO 465464

Mizuno, Hideyuki

Search repository
Mori, Shinichiro

× Mori, Shinichiro

WEKO 465465

Mori, Shinichiro

Search repository
草野 陽介

× 草野 陽介

WEKO 465466

en 草野 陽介

Search repository
上坂 紗紀

× 上坂 紗紀

WEKO 465467

en 上坂 紗紀

Search repository
矢島 佳央理

× 矢島 佳央理

WEKO 465468

en 矢島 佳央理

Search repository
熊谷 始紀

× 熊谷 始紀

WEKO 465469

en 熊谷 始紀

Search repository
水野 秀之

× 水野 秀之

WEKO 465470

en 水野 秀之

Search repository
森 慎一郎

× 森 慎一郎

WEKO 465471

en 森 慎一郎

Search repository
抄録
内容記述タイプ Abstract
内容記述 We evaluated the CT number accuracy in determining a CT calibration method for treatment planning by use of a 256 multi-slice CT (MSCT) scanner. An electron density phantom, which extends full length in the longitudinal direction, was scanned by the 256 MSCT scanner in a single rotation. We inserted four types of samples (air, 100 % ethanol, 40 wt% aqueous K2HPO4, and water) into the phantom. The regions of interest were set for all image slices, and the average CT number was calculated in the transverse and longitudinal sections. For the transverse image direction, the CT number became lower toward the center of the phantom with almost all samples. The causes of these phenomena might be attributed to the effects of scattered radiation and those of beam hardening due to the high-atomic-number material (40 wt% aqueous K2HPO4). However, the phenomena were too complicated to allow us to determine their causes only from the present studies. Meanwhile, for increasing the accuracy of the CT number calibration, a single sample should be inserted, and this step should be repeated with different samples. For the longitudinal direction, the CT number for a 40 wt % aqueous K2HPO4 sample increased by 30 HU from the anode to the cathode side due to variations in the X-ray energy caused by the heel effect. The caveat is that the CT number varies in the longitudinal direction when a CT scanner with a wide beam width is used.
書誌情報 Radiological Physics and Technology

巻 6, p. 241-247, 発行日 2013-02
ISSN
収録物識別子タイプ ISSN
収録物識別子 1865-0333
関連サイト
識別子タイプ URI
関連識別子 http://www.editorialmanager.com/rpte/
関連名称 http://www.editorialmanager.com/rpte/
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 23:48:52.047833
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3