{"_buckets": {"deposit": "7848285c-4a89-4402-a560-222aba58d324"}, "_deposit": {"created_by": 1, "id": "46555", "owners": [1], "pid": {"revision_id": 0, "type": "depid", "value": "46555"}, "status": "published"}, "_oai": {"id": "oai:repo.qst.go.jp:00046555", "sets": ["1"]}, "author_link": ["463993", "463986", "463989", "464000", "463994", "463991", "463997", "463999", "463987", "463985", "463984", "463998", "463990", "463995", "463996", "463992", "464001", "463988"], "item_8_biblio_info_7": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2013-05", "bibliographicIssueDateType": "Issued"}, "bibliographicPageEnd": "88", "bibliographicPageStart": "83", "bibliographicVolumeNumber": "723", "bibliographic_titles": [{"bibliographic_title": "Nuclear Instruments \u0026 Methods in Physics Research Section A"}]}]}, "item_8_description_5": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "The X\u0027tal cube is a depth-of-interaction (DOI)-PET detector which is aimed at obtaining isotropic resolution by effective readout of scintillation photons from six sides of the crystal block. The X\u0027tal cube is composed of a 3D crystal block with isotropic segments. Each face of the 3D crystal block is covered with a 4*4 array of multi-pixel photon counters (MPPCs). Previously, in order to fabricate the 3D crystal block efficiently and precisely, we applied a sub-surface laser engraving technique to a monolithic crystal block instead of gluing segmented small crystals. A dense arrangement of multiple micro-cracks carved by the laser beam works efficiently as a scattering wall for the scintillation photons. The X\u0027tal cube with the laser-processed block showed excellent performance with respect to crystal identification and energy resolution. In this work, for characteristics comparison between the laser-processed block and the conventional segmented array block, we made the laser-processed block and two types of segmented array blocks, one with air gaps and the other with glued segmented small crystals. All crystal blocks had 3D grids of 2 mm pitch. The 4*4 MPPC arrays were optically coupled to each surface of the crystal block. When performance was evaluated using a uniform irradiation of 511 keV, we found that the X\u0027tal cubes with the laser-processed block could easily achieve 2 mm3 uniform crystal identification. Also, the average energy resolution of each 3D grid was 11.1+-0.7%. On the other hand, the glued segmented array block had a pinched distribution and crystals could not be separated clearly. The segmented array block with air gaps had satisfactory crystal identification performance; however, the laser-processed block had higher crystal identification performance. Also, the energy resolution of the laser-processed block was better than for the segmented array blocks. In summary, we found the laser-processed X\u0027tal cube had excellent crystal identification performance and high energy performance due to its effective dense arrangement of micro-cracks.", "subitem_description_type": "Abstract"}]}, "item_8_relation_17": {"attribute_name": "関連サイト", "attribute_value_mlt": [{"subitem_relation_name": [{"subitem_relation_name_text": "http://dx.doi.org/10.1016/j.nima.2013.05.018"}], "subitem_relation_type_id": {"subitem_relation_type_id_text": "http://dx.doi.org/10.1016/j.nima.2013.05.018", "subitem_relation_type_select": "DOI"}}]}, "item_8_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "0168-9002", "subitem_source_identifier_type": "ISSN"}]}, "item_access_right": {"attribute_name": "アクセス権", "attribute_value_mlt": [{"subitem_access_right": "metadata only access", "subitem_access_right_uri": "http://purl.org/coar/access_right/c_14cb"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Yoshida, Eiji"}], "nameIdentifiers": [{"nameIdentifier": "463984", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Hirano, Yoshiyuki"}], "nameIdentifiers": [{"nameIdentifier": "463985", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Tashima, Hideaki"}], "nameIdentifiers": [{"nameIdentifier": "463986", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Inadama, Naoko"}], "nameIdentifiers": [{"nameIdentifier": "463987", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Nishikido, Fumihiko"}], "nameIdentifiers": [{"nameIdentifier": "463988", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Moriya, Takahiro"}], "nameIdentifiers": [{"nameIdentifier": "463989", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Omura, Tomohide"}], "nameIdentifiers": [{"nameIdentifier": "463990", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Watanabe, Mitsuo"}], "nameIdentifiers": [{"nameIdentifier": "463991", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Murayama, Hideo"}], "nameIdentifiers": [{"nameIdentifier": "463992", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Yamaya, Taiga"}], "nameIdentifiers": [{"nameIdentifier": "463993", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "吉田 英治", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "463994", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "平野 祥之", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "463995", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "田島 英朗", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "463996", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "稲玉 直子", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "463997", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "錦戸 文彦", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "463998", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "大村 知秀", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "463999", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "村山 秀雄", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "464000", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "山谷 泰賀", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "464001", "nameIdentifierScheme": "WEKO"}]}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "journal article", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "The X\u0027tal cube PET detector with a monolithic crystal processed by the 3D sub-surface laser engraving technique: Performance comparison with glued crystal elements", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "The X\u0027tal cube PET detector with a monolithic crystal processed by the 3D sub-surface laser engraving technique: Performance comparison with glued crystal elements"}]}, "item_type_id": "8", "owner": "1", "path": ["1"], "permalink_uri": "https://repo.qst.go.jp/records/46555", "pubdate": {"attribute_name": "公開日", "attribute_value": "2013-06-03"}, "publish_date": "2013-06-03", "publish_status": "0", "recid": "46555", "relation": {}, "relation_version_is_last": true, "title": ["The X\u0027tal cube PET detector with a monolithic crystal processed by the 3D sub-surface laser engraving technique: Performance comparison with glued crystal elements"], "weko_shared_id": -1}
The X'tal cube PET detector with a monolithic crystal processed by the 3D sub-surface laser engraving technique: Performance comparison with glued crystal elements
https://repo.qst.go.jp/records/46555
https://repo.qst.go.jp/records/4655520eb6094-3c15-47c5-b77d-ba883d002e15
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2013-06-03 | |||||
タイトル | ||||||
タイトル | The X'tal cube PET detector with a monolithic crystal processed by the 3D sub-surface laser engraving technique: Performance comparison with glued crystal elements | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Yoshida, Eiji
× Yoshida, Eiji× Hirano, Yoshiyuki× Tashima, Hideaki× Inadama, Naoko× Nishikido, Fumihiko× Moriya, Takahiro× Omura, Tomohide× Watanabe, Mitsuo× Murayama, Hideo× Yamaya, Taiga× 吉田 英治× 平野 祥之× 田島 英朗× 稲玉 直子× 錦戸 文彦× 大村 知秀× 村山 秀雄× 山谷 泰賀 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The X'tal cube is a depth-of-interaction (DOI)-PET detector which is aimed at obtaining isotropic resolution by effective readout of scintillation photons from six sides of the crystal block. The X'tal cube is composed of a 3D crystal block with isotropic segments. Each face of the 3D crystal block is covered with a 4*4 array of multi-pixel photon counters (MPPCs). Previously, in order to fabricate the 3D crystal block efficiently and precisely, we applied a sub-surface laser engraving technique to a monolithic crystal block instead of gluing segmented small crystals. A dense arrangement of multiple micro-cracks carved by the laser beam works efficiently as a scattering wall for the scintillation photons. The X'tal cube with the laser-processed block showed excellent performance with respect to crystal identification and energy resolution. In this work, for characteristics comparison between the laser-processed block and the conventional segmented array block, we made the laser-processed block and two types of segmented array blocks, one with air gaps and the other with glued segmented small crystals. All crystal blocks had 3D grids of 2 mm pitch. The 4*4 MPPC arrays were optically coupled to each surface of the crystal block. When performance was evaluated using a uniform irradiation of 511 keV, we found that the X'tal cubes with the laser-processed block could easily achieve 2 mm3 uniform crystal identification. Also, the average energy resolution of each 3D grid was 11.1+-0.7%. On the other hand, the glued segmented array block had a pinched distribution and crystals could not be separated clearly. The segmented array block with air gaps had satisfactory crystal identification performance; however, the laser-processed block had higher crystal identification performance. Also, the energy resolution of the laser-processed block was better than for the segmented array blocks. In summary, we found the laser-processed X'tal cube had excellent crystal identification performance and high energy performance due to its effective dense arrangement of micro-cracks. | |||||
書誌情報 |
Nuclear Instruments & Methods in Physics Research Section A 巻 723, p. 83-88, 発行日 2013-05 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0168-9002 | |||||
関連サイト | ||||||
識別子タイプ | DOI | |||||
関連識別子 | http://dx.doi.org/10.1016/j.nima.2013.05.018 | |||||
関連名称 | http://dx.doi.org/10.1016/j.nima.2013.05.018 |