量研学術機関リポジトリ「QST-Repository」は、国立研究開発法人 量子科学技術研究開発機構に所属する職員等が生み出した学術成果(学会誌発表論文、学会発表、研究開発報告書、特許等)を集積しインターネット上で広く公開するサービスです。 Welcome to QST-Repository where we accumulates and discloses the academic research results(Journal Publications, Conference presentation, Research and Development Report, Patent, etc.) of the members of National Institutes for Quantum and Radiological Science and Technology.
Thank you very much for using our website. On the 11th of March 2019, this site was moved from our own network server to the JAIRO Cloud network server. If you previously bookmarked this site, that bookmark will no longer work. We would be grateful if you could bookmark the website again. Thank you very much for your understanding and cooperation.
Intraoperative imaging has been studied using conventional devices such as near infrared (NIR) optical probes and gamma probes. However, these devices have limited depth penetration and spatial resolution. In a previous study, we realized a multi-modal endoscopic system. However, charge-coupled device (CCD)-based gamma imaging required long acquisition times and lacked gamma energy information. A silicon photomultiplier (SiPM)-based gamma detector is implemented in a multi-modal laparoscope herein. A gradient index (GRIN) lens and CCD are used to transfer and readout visible and NIR photons. The feasibility of in-vivo sentinel lymph node (SLN) mapping was successfully performed with the proposed system.