量研学術機関リポジトリ「QST-Repository」は、国立研究開発法人 量子科学技術研究開発機構に所属する職員等が生み出した学術成果(学会誌発表論文、学会発表、研究開発報告書、特許等)を集積しインターネット上で広く公開するサービスです。 Welcome to QST-Repository where we accumulates and discloses the academic research results(Journal Publications, Conference presentation, Research and Development Report, Patent, etc.) of the members of National Institutes for Quantum Science and Technology.
Thank you very much for using our website. On the 11th of March 2019, this site was moved from our own network server to the JAIRO Cloud network server. If you previously bookmarked this site, that bookmark will no longer work. We would be grateful if you could bookmark the website again. Thank you very much for your understanding and cooperation.
Background
For performing a quantitative EPR measurement, an external and/or internal standard sample having a known spin concentration is required. If adequate reproducibility of the sample volume, shape, and positioning in the cavity for every EPR measurement can be guaranteed, the external standard and the sample tested can be directly compared and then the process of comparisons for quantification can be simplified.
Methods
0.1 mM TEMPOL sample as a standard sample loaded into the capillary was measured with a X-band EPR (JEOL, Tokyo, Japan).
Reproducibilities of EPR signal intensity of the 0.1 mM TEMPOL standard were estimated for three different types of sample tubes; a PTFE tubing, a glass capillary, and a quartz flat cuvette. A special silicon cap was put on the top of the sample holder to fix the glass capillary. Again, the reproducibility was tested with or without the silicon cap.
Results
The cuvette could show the most suitable reproducibility among the three sample tubes. Reproducibility and sensitivity of the EPR signal intensity was improved when the standard sample in a capillary was measured using the silicon cap equipment.
Conclusion
Suppressing the sample movement could improve reproducibility and sensitivity of the EPR spectroscopy. For a quantitative EPR measurement, retaining the volume and position of the sample in the cavity is quite important.
会議概要(会議名, 開催地, 会期, 主催者等)
9th Biennial Meeting for the Society for Free Radical Research Asia