量研学術機関リポジトリ「QST-Repository」は、国立研究開発法人 量子科学技術研究開発機構に所属する職員等が生み出した学術成果(学会誌発表論文、学会発表、研究開発報告書、特許等)を集積しインターネット上で広く公開するサービスです。 Welcome to QST-Repository where we accumulates and discloses the academic research results(Journal Publications, Conference presentation, Research and Development Report, Patent, etc.) of the members of National Institutes for Quantum and Radiological Science and Technology.
Thank you very much for using our website. On the 11th of March 2019, this site was moved from our own network server to the JAIRO Cloud network server. If you previously bookmarked this site, that bookmark will no longer work. We would be grateful if you could bookmark the website again. Thank you very much for your understanding and cooperation.
The traceability plays an important role in the quality control systems for any measurement. The desire to improve of it as well as harmonize of radon and thoron measurements in air around the world has been lastly increasing because radon and thoron activity concentrations in air measured with different radon and thoron monitors in the same environment were found to be inconsistent with each other.
\nThe aim of this work is to validate the traceability of the existing radon and thoron facilities located in four Asian countries (China, India, Japan, Thailand) for values ranging from 500 Bq/m3 to 5000 Bq/m3 and 4000 Bq/m3 to 10 000 Bq/m3, for radon and thoron, respectively.
\nIn this project, a comparison among five facilities is performed using one reference device for radon and one for thoron. The radon reference device calibrated with the secondary radon standard traceable to PTB (Germany), NIST (USA) and NPL (UK) to be exposed in participants’ radon facilities for a comparison with their reference monitor. The thoron passive reference device have been checked in the NIRS thoron chamber, and is exposed in the participants’ thoron chambers for a comparison with their reference devices.
\nAs a result, a comparison reference value and its uncertainty have been derived for each activity concentration. Finally, bias and degrees of equivalence between laboratories have been statistically analysed and presented.
会議概要(会議名, 開催地, 会期, 主催者等)
AOCRP 5 (5th Asian & Oceanic IRPA Regional Congress on Radiation Protection) における発表