量研学術機関リポジトリ「QST-Repository」は、国立研究開発法人 量子科学技術研究開発機構に所属する職員等が生み出した学術成果(学会誌発表論文、学会発表、研究開発報告書、特許等)を集積しインターネット上で広く公開するサービスです。 Welcome to QST-Repository where we accumulates and discloses the academic research results(Journal Publications, Conference presentation, Research and Development Report, Patent, etc.) of the members of National Institutes for Quantum Science and Technology.
Thank you very much for using our website. On the 11th of March 2019, this site was moved from our own network server to the JAIRO Cloud network server. If you previously bookmarked this site, that bookmark will no longer work. We would be grateful if you could bookmark the website again. Thank you very much for your understanding and cooperation.
Current clinical PET scanners have a 15–22 cm axial field-of-view (FOV). These scanners image the whole body using six or more bed positions. We designed an axially extendable multiplex cylinder (AEMC) PET scanner to provide high versatility for clinical and research studies using semiconductor photo-sensor based, depth-of-interaction (DOI) detectors. Since silicon-photomultipliers (Si-PMs) have high gain like conventional photomultiplier tubes and a compact design, the Si-PM-based detector is particularly expected to enable various new detector arrangements. The AEMC-PET scanner consists of multiple independent and laminated detector rings using the DOI detectors. The AEMC-PET scanner can extend the axial FOV as each stacked detector ring can be slid sideways. When this PET scanner is used for the four-layer DOI detector, its minimum axial FOV is 24 cm and its maximum crystal thickness is 3 cm. On the other hand, the axial FOV can be extended to 96 cm when laminated detector rings are slid sideways, but the crystal thickness must be 1/4 of 3 cm. In this work, we evaluated performance characteristics of the PET scanner with a variable axial FOV using Monte Carlo simulation. From the simulation of the 180-cm line source, the 96-cm axial FOV was found to have two-fold better sensitivity compared to the 24-cm axial FOV. For extension of the axial FOV, scatter and attenuation of oblique lines-of-response reduced the yield of true coincidences, but effects of scatter and attenuation were small. Conclusive results were obtained showing the 52.8-cm axial FOV yielded an increase in the noise equivalent count rate of approximately 30% relative to the 24-cm axial FOV. We expect the designed AEMC-PET scanner will provide high versatility in applications such as for measuring whole-body tracer uptakes while keeping the continuous axial FOV; as well, the scan time for static images will be reduced for a comparable number of detectors as conventional PET scanners.