量研学術機関リポジトリ「QST-Repository」は、国立研究開発法人 量子科学技術研究開発機構に所属する職員等が生み出した学術成果(学会誌発表論文、学会発表、研究開発報告書、特許等)を集積しインターネット上で広く公開するサービスです。 Welcome to QST-Repository where we accumulates and discloses the academic research results(Journal Publications, Conference presentation, Research and Development Report, Patent, etc.) of the members of National Institutes for Quantum Science and Technology.
Thank you very much for using our website. On the 11th of March 2019, this site was moved from our own network server to the JAIRO Cloud network server. If you previously bookmarked this site, that bookmark will no longer work. We would be grateful if you could bookmark the website again. Thank you very much for your understanding and cooperation.
A compact ECR ion source has been developed for heavy-ion medical facilities. The beam intensity and stability were considerably improved by recent modifications on three points (length of sextupole, cooling system for the extraction electrode and position of Einzel lens). Initial results of C4+ beam tests show that an intensity of 180 emA can be routinely obtained with simple tuning. The best record was 220 emA for C4+, which meets the medical requirements. Throughout these tests, CH4 gas was used with 0.1 cc/min and the extraction voltage was fixed at 25 kV. Results on beam emittance and long-term stability are also briefly discussed.